Keysight 41800A Active Probe
41
Performance Tests
Probe Gain/Frequency Response Tests
19.Press the
MKR
&
key and the
'MKR
&
MAX'
softkey to move the
*
marker to
the maximum point between 200 MHz and 500 MHz, and confirm that the
data T/R is 1.5 dB/–2 dB.
20.Press the
'MKR
&
MIN'
softkey to move the
*
marker to the minimum point
between 200 MHz and 500 MHz, and confirm that the T/R data is within
+1.5 dB/–2 dB.