Keysight 41800A Active Probe
57
Service
Disassembly
7. Draw the coaxial cable and leads, a part of cable assembly PN
41800-61601, from the CASE toward the arrow A as shown in
8. Remove the solder at point 9 as shown in
.
9. Remove the N(m) connector from the CASE toward the arrow B shown
, while holding the soldering iron at point 9 as shown in
10.Remove the four screws holding the Amplifier/Regulator Board assembly in
place.
11.Remove the Amplifier/Regulator Board assembly from the CASE.