Keysight 41800A Active Probe
21
Operation
Preparation for Use
Procedure
1. Connect the equipment as shown in
2. Set up the Keysight 4195A as follows.
3. Confirm that the signal level is
–20 dBm ± 5 dB
.
Operating Check Using a Network Analyzer
shows the operating check setup using a network analyzer. To
check the probe gain, absolute power amplitude measurement mode should
be selected.
Figure 3-4
Operating Check Setup Using a Network Analyzer
CONFIG
SPECTRUM
PORT SELECT
SOURCE CH1 on
CENTER
50 MHz
SPAN
0 Hz
AMPLITUDE (S1)
–26 dBm
REF ATTEN (R1)
10 dB
The Keysight 11880A probe-N(m) adapter can be used instead of the
N(m)-BNC(f) adapter and the Keysight 10218A probe-BNC(m) adapter. The
signal level should be –26 dBm ± 5 dB when the Keysight 11880A
probe-N(m) adapter is used.