Keysight 41800A Active Probe
Performance Tests
Probe Gain/Frequency Response Tests
4-
Equipment
Procedure
1. Connect the power splitter to the Keysight 4195A as shown in
(1).
2. Sequentially press the following Keysight 4195A front panel keys, to set the
Keysight 4195A to the Probe Gain/Frequency Response test settings.
CONFIG, 'NETWORK', PRESET
AMPLITUDE
(CHANNEL 1 side)
, -, 1, 4, ENTER/EXECUTE
REF ATTEN
(CHANNEL 1 side)
, 0, ENTER/EXECUTE
TEST ATTEN
(CHANNEL 1 side)
, 0, ENTER/EXECUTE
START, 1, 0, Hz/dBµV
STOP, 5, 0, 0, MHz/V
MENU, 'TRIGGER menu', 'SINGLE mode', 'return'
'RESOLUTN menu', 'No. of POINTS', 1, 0, 1, ENTER/EXECUTE
'return', 'TYPE lin log'
(
“log”
in the
'TYPE lin log'
softkey
will change to intensified green.)
DISPLAY, 'TRACE B on off'
(
“off”
in the
'TRACE B on off'
softkey
will change to intensified green.)
AUTO
(The yellow LED in the
AUTO
key will
turn on.)
3. Sequentially press the following Keysight 4195A front panel keys to set the
Keysight 4195A's programmed points table, and to activate its program
sweep measurement capability.
MENU, 'PROGRAM sweep', 'PROG TBL set up'
(Confirm that the displayed table is empty. If the table is not empty, press
the
'TABLE No.'
softkey repeatedly until an empty table is displayed, or
press the
'TABLE ALL CLR'
softkey and
ENTER/EXECUTE
key.)
Network Analyzer
Keysight 4195A
Power Splitter
Keysight 11667A
Probe to N(m) Adapter Keysight 11880A
N(m)-N(m) Cable
Keysight 11500B
N(f)-N(f) Adapter
PN 1250-0777
N(m)-N(m) Adapter
PN 1250-0778