Keysight 41800A Active Probe
49
Adjustment
LF Gain/Flatness Adjustment
Procedure
1. Connect the power splitter to the Keysight 4195A as shown in
(1).
2. Sequentially press the following Keysight 4195A front panel keys, to set the
Keysight 4195A to the adjustment settings.
CONFIG, 'NETWORK', PRESET
AMPLITUDE
(CHANNEL 1 side)
, -, 1, 4, ENTER/EXECUTE
REF ATTEN
(CHANNEL 1 side)
, 0, ENTER/EXECUTE
TEST ATTEN
(CHANNEL 1 side)
, 0, ENTER/EXECUTE
START, 1, 0, Hz/dBµV
STOP, 5, 0, MHz/V
MENU, 'TRIGGER menu', 'SINGLE mode', 'return'
'RESOLUTN menu', 'No. of POINTS', 2, 1, ENTER/EXECUTE
'return', 'TYPE lin log'
(
“log”
in the
'TYPE lin log'
softkey
will change to intensified green.)
DISPLAY, 'TRACE B on off'
(
“off”
in the
'TRACE B on off'
softkey
will change to intensified green.)
AUTO
(The yellow LED in the
AUTO
key will
turn on.)
3. Sequentially press the following keys to perform a normalize (thorough)
calibration, and wait until the message “Cal completed” is displayed.
CAL, 'TRANS CAL menu', 'NORMLIZE (THRU)', 'THRU', ENTER/EXECUTE
4. Press the Keysight 4195A's
CAL
key and the
'CORRECTN on off'
softkey to
activate the Keysight 4195A's correction capability. The “
on
” in this softkey
will change to intensified green.
(1)) from the power splitter, and
connect the 41800A between the power splitter and CABLE 1 as shown in
(2).
6. Press the Keysight 4195A's
TRIG/RESET
key to make a single sweep
measurement.
7. Enter the “REF=A(21)+0.5” command and “DIV=0.1” command to set the
scale of the Keysight 4195A's display, using the following key sequence.
SCALE REF, 'A REF LEVEL',
blue shiftkey
, A, MATH OPERATOR,
'(', 2, 1, ')', ' + ', 0, ., 5, ENTER/EXECUTE
SCALE REF, 'A /DIV', 0, ., 1, ENTER/EXECUTE