71
Chapter 3 Features and Functions
Pulse Waveforms
3
Edge Time
The edge time sets each edge transition time (rising and falling) of the
pulse. The rise time and fall time cannot be set independently—each is
equal to the edge time. For each transition, the edge time represents the
time from the 10% threshold to the 90% threshold.
• Edge time: 20 ns to 100 ns (see restrictions below). The default edge
time is 20 ns.
• The specified edge time must fit within the specified pulse width as
shown below. The function generator will adjust the edge time as
needed to accommodate the specified pulse width.
Edge Time < 0.625
X
Pulse Width
or in terms of duty cycle
Edge Time < 0.625
X
Period
X
Duty Cycle ÷ 100
• Front-Panel Operation: After selecting the pulse function, press the
Edge Time
softkey. Then use the knob or numeric keypad to enter the
desired edge time.
• Remote Interface Operation:
FUNCtion:PULSe:TRANsition {<
seconds
>|MINimum|MAXimum}
33210A users guide.book Page 71 Wednesday, July 16, 2008 11:16 AM
Summary of Contents for 33210A
Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator User s Guide ...
Page 2: ......
Page 3: ......
Page 14: ...12 Contents Contents 33210A users guide book Page 12 Wednesday July 16 2008 11 16 AM ...
Page 15: ...1 1 Quick Start 33210A users guide book Page 13 Wednesday July 16 2008 11 16 AM ...
Page 53: ...3 3 Features and Functions 33210A users guide book Page 51 Wednesday July 16 2008 11 16 AM ...
Page 273: ...5 5 Error Messages 33210A users guide book Page 271 Wednesday July 16 2008 11 16 AM ...
Page 301: ...6 6 Application Programs 33210A users guide book Page 299 Wednesday July 16 2008 11 16 AM ...
Page 311: ...7 7 Tutorial 33210A users guide book Page 309 Wednesday July 16 2008 11 16 AM ...
Page 335: ...8 8 Specifications 33210A users guide book Page 333 Wednesday July 16 2008 11 16 AM ...