138
Chapter 3 Features and Functions
External Timebase Reference (Option 001)
3
4. Leaving the phase setting at its default (zero) on the first 33210A, use
the
Adjust Phase
feature to adjust the phase of the second 33210A to
align the output signals:
You can use the
Set 0 Phase
feature to set a new zero-phase reference
point with the two function generators now aligned.
• Front-Panel Operation: Press
and select the
Output Setup,
softkey, then
Adjust Phase
. Set the phase angle (±360 degrees) using
the knob and cursor keys or the numeric keypad. Once you have
aligned the signals on the scope, press
Set 0 Phase
to set a new zero-
phase reference point.
• Remote Interface Operation: Use the following command to set the
phase offset (the query returns the currently set offset):
PHASe {<
angle
>|MINimum|MAXimum}
PHASe? [MINimum|MAXimum]
• The following command sets a new zero-phase reference point:
PHASe:REFerence
Application Note 1426:
“How to Connect Two or More Signal
Generators to Create a Multi-Channel Waveform Generator” provides
additional information on this subject.
Out of Phase
Aligned
33210A users guide.book Page 138 Wednesday, July 16, 2008 11:16 AM
Summary of Contents for 33210A
Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator User s Guide ...
Page 2: ......
Page 3: ......
Page 14: ...12 Contents Contents 33210A users guide book Page 12 Wednesday July 16 2008 11 16 AM ...
Page 15: ...1 1 Quick Start 33210A users guide book Page 13 Wednesday July 16 2008 11 16 AM ...
Page 53: ...3 3 Features and Functions 33210A users guide book Page 51 Wednesday July 16 2008 11 16 AM ...
Page 273: ...5 5 Error Messages 33210A users guide book Page 271 Wednesday July 16 2008 11 16 AM ...
Page 301: ...6 6 Application Programs 33210A users guide book Page 299 Wednesday July 16 2008 11 16 AM ...
Page 311: ...7 7 Tutorial 33210A users guide book Page 309 Wednesday July 16 2008 11 16 AM ...
Page 335: ...8 8 Specifications 33210A users guide book Page 333 Wednesday July 16 2008 11 16 AM ...