177
Chapter 4 Remote Interface Reference
Output Configuration Commands
4
• You can also set the offset by specifying a high level and low level.
For example, if you set the high level to +2 volts and the low level to
-3 volts, the resulting amplitude is 5 Vpp (with an associated offset
voltage of -500 mV). See the
VOLT:HIGH
and
VOLT:LOW
commands
below for more information.
• To output a dc voltage level, select the dc voltage function using the
FUNC DC
command and then set the offset voltage level using the
VOLT:OFFS
command. You can set the dc level to any value between
±
5 Vdc into 50 ohms or
±
10 Vdc into an open circuit.
VOLTage
:HIGH
{<
voltage
>|
MINimum
|
MAXimum
}
:HIGH?
[
MINimum
|
MAXimum
]
:LOW
{<
voltage
>|
MINimum
|
MAXimum
}
:LOW?
[
MINimum
|
MAXimum
]
Set the high or low voltage levels. For all functions, the default high level
is +50 mV and the default low level is -50 mV. MIN selects the most
negative voltage level for the selected function and MAX selects the
largest voltage level. The
:HIGH?
and
:LOW?
queries return the high and
low levels, respectively.
• Limits Due to Amplitude: You can set the voltage levels to a positive
or negative value with the restrictions shown below. Vpp is the
maximum peak-to-peak amplitude for the selected output termination
(10 Vpp for a 50-ohm load or 20 Vpp for a high-impedance load).
Vhigh – Vlow < Vpp (max)
and
Vhigh, Vlow <
If the specified level is not valid, the function generator will
automatically adjust it to the maximum voltage allowed. From the
remote interface, a “Data out of range” error will be generated and
the level will be adjusted as described.
• You can set the levels to a positive or negative value, but note that
the high level must always be greater than the low level. If you
specify a low level that is greater than the high level, the function
generator will automatically set the high level to be 1 mV more than
the new low level. A “Settings conflict” error will be generated.
Vpp max
(
)
2
-------------------------------
33210A users guide.book Page 177 Wednesday, July 16, 2008 11:16 AM
Summary of Contents for 33210A
Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator User s Guide ...
Page 2: ......
Page 3: ......
Page 14: ...12 Contents Contents 33210A users guide book Page 12 Wednesday July 16 2008 11 16 AM ...
Page 15: ...1 1 Quick Start 33210A users guide book Page 13 Wednesday July 16 2008 11 16 AM ...
Page 53: ...3 3 Features and Functions 33210A users guide book Page 51 Wednesday July 16 2008 11 16 AM ...
Page 273: ...5 5 Error Messages 33210A users guide book Page 271 Wednesday July 16 2008 11 16 AM ...
Page 301: ...6 6 Application Programs 33210A users guide book Page 299 Wednesday July 16 2008 11 16 AM ...
Page 311: ...7 7 Tutorial 33210A users guide book Page 309 Wednesday July 16 2008 11 16 AM ...
Page 335: ...8 8 Specifications 33210A users guide book Page 333 Wednesday July 16 2008 11 16 AM ...