137
Chapter 3 Features and Functions
External Timebase Reference (Option 001)
3
External Timebase Reference (Option 001)
Option 001 is the External Timebase Reference, which provides rear
panel connectors (
10 MHz In
and
10 MHz Out
) and circuitry to allow
synchronization between multiple Keysight 33210A function generators
or to an external 10 MHz clock signal. You can also set the phase offset
of the output waveform from the front panel or over the remote
interface. To synchronize multiple 33210As,
all
of them must have
Option 001 installed.
To align the phase of two 33210As, you can use a dual-channel
oscilloscope to compare the output signals:
1. Connect two 33210As, with
10 MHz Out
connected to
10 MHz In
:
2. Connect the outputs of the 33210As to the scope channel inputs:
3. Set the same frequency on both function generators. The scope should
show the signals to be in sync with regard frequency, but not phase.
(A square wave works well to show the phase difference.)
33210A users guide.book Page 137 Wednesday, July 16, 2008 11:16 AM
Summary of Contents for 33210A
Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator User s Guide ...
Page 2: ......
Page 3: ......
Page 14: ...12 Contents Contents 33210A users guide book Page 12 Wednesday July 16 2008 11 16 AM ...
Page 15: ...1 1 Quick Start 33210A users guide book Page 13 Wednesday July 16 2008 11 16 AM ...
Page 53: ...3 3 Features and Functions 33210A users guide book Page 51 Wednesday July 16 2008 11 16 AM ...
Page 273: ...5 5 Error Messages 33210A users guide book Page 271 Wednesday July 16 2008 11 16 AM ...
Page 301: ...6 6 Application Programs 33210A users guide book Page 299 Wednesday July 16 2008 11 16 AM ...
Page 311: ...7 7 Tutorial 33210A users guide book Page 309 Wednesday July 16 2008 11 16 AM ...
Page 335: ...8 8 Specifications 33210A users guide book Page 333 Wednesday July 16 2008 11 16 AM ...