116
Chapter 3 Features and Functions
Optional Arbitrary Waveforms (Option 002)
3
10 Define the remaining waveform points.
Using the
Time
and
Voltage
softkeys, define the remaining waveform
points using the values shown in the table below.
• The time value of the last point that can be defined in the waveform
must be less than the specified cycle period.
• The waveform editor automatically connects the last waveform point
to the voltage level of first point to create a continuous waveform.
• To insert additional points after the current waveform point, press
the
Insert Point
softkey. The new point is inserted midway between
the current point and the next defined point.
• To remove the current waveform point, press the
Remove Point
softkey. The remaining points are joined using the interpolation
method currently selected. You cannot remove Point #1 since the
waveform must have a defined initial value.
Point
Time Value
Voltage Value
1
0 s
0 V
2
2 ms
3 V
3
4 ms
1 V
4
7 ms
0 V
33210A users guide.book Page 116 Wednesday, July 16, 2008 11:16 AM
Summary of Contents for 33210A
Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator User s Guide ...
Page 2: ......
Page 3: ......
Page 14: ...12 Contents Contents 33210A users guide book Page 12 Wednesday July 16 2008 11 16 AM ...
Page 15: ...1 1 Quick Start 33210A users guide book Page 13 Wednesday July 16 2008 11 16 AM ...
Page 53: ...3 3 Features and Functions 33210A users guide book Page 51 Wednesday July 16 2008 11 16 AM ...
Page 273: ...5 5 Error Messages 33210A users guide book Page 271 Wednesday July 16 2008 11 16 AM ...
Page 301: ...6 6 Application Programs 33210A users guide book Page 299 Wednesday July 16 2008 11 16 AM ...
Page 311: ...7 7 Tutorial 33210A users guide book Page 309 Wednesday July 16 2008 11 16 AM ...
Page 335: ...8 8 Specifications 33210A users guide book Page 333 Wednesday July 16 2008 11 16 AM ...