Key Reference
BW/SWP
86
BW/SWP
Activates the bandwidth function and accesses the
associated softkeys to control resolution bandwidth,
video bandwidth and sweep time.
RBW
Key access:
[BW/SWP] > {RBW}
Changes the 3 dB resolution bandwidth on the analyzer
from 30 Hz to 1 MHz in a 1, 3, 10 sequence using the
knob, step keys or the numeric keypad.
As the resolution bandwidth is decreased, the sweep
time is modified to maintain amplitude calibration.
Resolution bandwidth is also related to span. As span is
decreased, resolution bandwidth decreases. As the
resolution bandwidth changes, the video bandwidth, if
in auto couple mode, changes to maintain the
VBW/RBW ratio.
A “#”mark appears next to RBW on the display when it is
not coupled.
VBW
Key access:
[BW/SWP] > {VBW}
Changes the analyzer post-detector filter from 3 Hz to 1
MHz in a 1, 3, 10 sequence using the knob, the step
keys, or the numeric keypad.
As the video bandwidth is decreased, the sweep time is
increased to maintain amplitude calibration.
NOTE
If an unavailable bandwidth is entered using the numeric keypad,
the closest available bandwidth in the 1, 3, 10 sequence is used.
NOTE
In non-zero span mode, the value of RBW is determined by
whether the setting for RBW is Auto or Man (manual).
NOTE
If an unavailable bandwidth entered using the numeric keypad,
the closest available bandwidth in the 1, 3, 10 sequence is used.
Summary of Contents for N9340B
Page 7: ...1 Overview 1 Overview ...
Page 42: ...System and File File 36 ...
Page 43: ...37 Making Measurements 4 Making Measurements ...
Page 150: ...Error Messages Error Message List 144 ...
Page 154: ...Troubleshooting Contact Keysight Technologies 148 Contact Keysight Technologies ...
Page 157: ...Menu Map AMPTD 151 ...