Key Reference
MODE
113
Amptd Step
Key access:
[MODE] > {Tracking Generator } > { Amptd
Step}
Allows you to set the step size of the power level range
of the tracking generator. The default setting is one
vertical scale division when in a logarithmic amplitude
scale.
Amptd Offset
Key access:
[MODE]>{Tracking Generator}> { Amptd
Offset}
Offsets the displayed power of the tracking generator.
Using the amplitude offset capability of the tracking
generator allows you to take system losses or gains into
account, thereby displaying the actual power delivered
to the device under test.
CAU-
CAUTION
Power-level sensitive devices connected to the tracking
generator output may be accidentally damaged. This is because
the actual source amplitude will be greater than the amplitude
indicated on the analyzer, when the power sweep function is
being used. The source amplitude value is used as the starting
amplitude for a power sweep. The ending amplitude for a power
sweep is higher then the source amplitude value and it may
actually be set to values beyond specified output levels.
Therefore, exercise caution when connecting a power-level
sensitive device to the tracking generator output.
NOTE
When Amplitude (Off) is selected, the output attenuation on the
N9340B is set to the maximum attenuation.
Summary of Contents for N9340B
Page 7: ...1 Overview 1 Overview ...
Page 42: ...System and File File 36 ...
Page 43: ...37 Making Measurements 4 Making Measurements ...
Page 150: ...Error Messages Error Message List 144 ...
Page 154: ...Troubleshooting Contact Keysight Technologies 148 Contact Keysight Technologies ...
Page 157: ...Menu Map AMPTD 151 ...