Getting Started
Powering on the Analyzer for the First Time
18
Powering on the Analyzer for the First Time
Insert the battery into the analyzer or connect the
analyzer to an external line power supply via the AC-DC
adapter, then press the power switch on the front panel
of your N9340B to power on the analyzer.
Allow the analyzer to warm-up for 30 minutes before
making a calibrated measurement. To meet its
specifications, the analyzer must meet operating
temperature conditions.
CAU-
CAUTION
Use only the original AC-DC adapter or originally supplied
battery for the power source.
The maximum RF input level of an average continuous power
is 33 dBm (or +50 VDC signal input). Avoid connecting a
signal into the analyzer that exceeds the maximum level.
Install battery
Press Power Switch