Key Reference
Marker
99
Marker Trace
Key access:
[MARKER] > {More} > {Marker Trace}
Activates a marker on the trace if there are no markers
turned on. If a marker is currently active, press
Marker
Trace
until
1
,
2
,
3
, or
4
is underlined. The active marker
will be moved to the selected trace.
Selecting
Auto
will move the marker to the automatically
selected trace. The selection order is to look for the
lowest numbered trace in the following order of trace
modes:
If there is no trace in any of these modes, it selects trace
1 as default.
All Off
Key access:
[MARKER] > {More} > {All Off}
Turns off all of the markers, including markers used for
signal track and demodulation. The marker annotation is
also removed.
Marker Table
Key access:
[MARKER] > {More} > {Marker Table}
Compresses the graticule and displays the marker
number, trace number, marker type, X axis value, and
the amplitude in the table.
NOTE
Frequency counter is properly effective only on CW signals or
discrete peaks. For a valid reading, the marker must be greater
than 25 dB above the noise.
Freq Count
function is not
supported under
Marker Delta
condition.
NOTE
Noise
mode operates only on noise peaks and not supported
under
Marker Delta
condition.
Clear write
Max hold
Min. hold
View
Summary of Contents for N9340B
Page 7: ...1 Overview 1 Overview ...
Page 42: ...System and File File 36 ...
Page 43: ...37 Making Measurements 4 Making Measurements ...
Page 150: ...Error Messages Error Message List 144 ...
Page 154: ...Troubleshooting Contact Keysight Technologies 148 Contact Keysight Technologies ...
Page 157: ...Menu Map AMPTD 151 ...