Key Reference
Amplitude
83
Preamp
Key access:
[AMPTD] > {Preamp}
Toggles the internal preamp between
On
and
Off
.
Preamp results in a correction being applied to
compensate for the gain of the preamp at 20 dBm so
that amplitude readings show the actual value at the
input connector.
The preamp is set to on in frequency bands from
1 MHz to 3 GHz, otherwise the correction is not applied
even though the
PA
indication still appears.
Scale/DIV
Key access:
[AMPTD]
Sets the logarithmic units per vertical graticule division
on the display.
Scale Type
Key access:
[AMPTD] > {Scale Type}
Toggles the scale type of vertical graticule divisions
between logarithmic and linear.
When Scale Type is set to
Log
, a logarithmic units is
activated ranging from 1 to 10 dB per division.
When Scale Type is set to
Lin
, the vertical divisions are
linearly scaled and the default amplitude units are volts.
The top of the display is set to the reference-level value
and the bottom graticule becomes zero volts. (Each
division of the graticule is one-tenth of the reference
level in volts.)
NOTE
When the preamp is on, a PA indication appears on the left side of
the display.
NOTE
The Scale/Div function is only available when the Scale Type key is
set to Log.
Summary of Contents for N9340B
Page 7: ...1 Overview 1 Overview ...
Page 42: ...System and File File 36 ...
Page 43: ...37 Making Measurements 4 Making Measurements ...
Page 150: ...Error Messages Error Message List 144 ...
Page 154: ...Troubleshooting Contact Keysight Technologies 148 Contact Keysight Technologies ...
Page 157: ...Menu Map AMPTD 151 ...