102 N5511A Phase Noise Test System User’s Guide
Absolute Measurement Examples
Single Channel Measurement
Configuring Equipment
Configure the DUT and source for this measurement:
1) Set up DUT PSG
Frequency: 1
GHz
Amplitude: 11
dBm
RF On/Off: RF ON
Mod On/Off: MOD OFF
2) Set up REF PSG
Frequency: 1
GHz
Amplitude: 14
dBm
RF On/Off: RF ON
Mod On/Off: MOD OFF
The RF power of the DUT source must be at least 3 dB lower than that of the REF source