244 N5511A Phase Noise Test System User’s Guide
FM Discriminator Measurement Examples
FM Discriminator Single Channel Measurement using FM Rate and Deviation
Calibration
Cal Configuration
1.
2.
Select
Derive constant from FM rate and deviation
as the calibration
method.
3.
Set the
FM Rate
to 20 kHz and
FM Deviation
to 10 kHz, which are the
recommended FM rate and deviation.
Figure 10-14
Enter Parameters in the Cal Tab