104 N5511A Phase Noise Test System User’s Guide
Absolute Measurement Examples
Single Channel Measurement
1) Choose Measurement Type
Measurement Type: Absolute Phase
Noise (using phase locked loop)
2) Choose Measurement
Method
Flexible is "single segment"
measurement method where every
segment is captured independently
and after the previous segment has
been measured (this is how E5500
works).
Fast is "multi-segment", where
segments are being measured in
parallel. This option allows for a live
view of the data acquisition of the
measurement.
Measurement Method: Flexible
3) Choose Channel Setup
Channel Setup: Single
4) Set Start and Stop Offsets
Start Offset: 10 Hz
Stop Offset: 2 MHz
Summary of Contents for N5511A
Page 1: ...User s Guide Keysight N5511A Phase Noise Test System...
Page 4: ...4...
Page 11: ...Contents N5511A Phase Noise Test System User s Guide 11...
Page 12: ...12 N5511A Phase Noise Test System User s Guide Contents...
Page 38: ...38 N5511A Phase Noise Test System User s Guide Phase Noise Basics What is Phase Noise...