N5511A Phase Noise Test System User’s Guide 101
Absolute Measurement Examples
Single Channel Measurement
Test Setup
shows the configuration used for the Single Channel measurement.
Figure 6-1
Single Channel Measurement Setup
Connectors on M9550A
Connects to
Green
EFC Out on rear of PSG
Purple
Scope Monitor
Red
DUT source RF Out
Blue
Reference Source RF out
1)
CH 1 PLL Tune Line
Connect the Ch 1 PLL Tune Voltage Out port (green) of the M9550A phase
detector module inside the PNTS chassis to the EFC port at the back of the REF
(PSG as reference) source using the BNC-to-BNC cable.
2)
DUT to Sig/RF on PNTS
Connect the RF output of the DUT source to the high-band RF input (1.2 to
40 GHz) labeled "Sig/RF" on the M9550 phase detector module using an SMA
cable
3)
REF to Ref/LO on PNTS
Connect the RF output port of the REF source to the high-band RF input (1.2 to
40 GHz) labeled "Ref/LO" on the M9550 phase detector module (blue) using an
SMA cable
4)
MONITOR output to Scope
Connect SMB-to-BNC cable from the "Monitor" output on the M9550 phase
detector (purple) to channel 1 of the oscilloscope