N5511A Phase Noise Test System User’s Guide 363
Preventive Maintenance
General Procedures and Techniques
3.5 mm connectors with a silver hex nut
All other semirigid cable connectors use a silver-colored nut that can be
turned. To remove this type of connector, turn the silver nut counter-clockwise
with a 5/16 inch wrench.
When reconnecting this type of cable:
— Carefully insert the male connector center pin into the female connector.
(Make sure the cable is aligned with the instrument connector properly
before joining them.)
— Turn the silver nut clockwise by hand until it is snug, then tighten with an
8 inch-lb torque wrench (part number 8720-1765).
Bent Semirigid Cables
Semirigid cables are not intended to be bent outside of the factory. An
accidental bend that is slight or gradual may be straightened carefully by
hand. Semirigid cables that are crimped will affect system performance and
must be replaced. Do not attempt to straighten a crimped semirigid cable.