200 N5511A Phase Noise Test System User’s Guide
Residual Measurement Examples
Amplifier Measurement Example - Single Channel
Making the Measurement
Calibrate the measurement using Single-Sided Spur calibration.
Refer to
Chapter 7, “Residual and Additive Noise Measurement
for more information about residual phase noise
measurements calibration types.
If the software prompts to
Establish Quadrature Manually
adjust the phase
difference at the phase detector until the voltmeter at the right-hand side of
the display, shown below, reads approximately 0 V. Click
Continue
.
Figure 8-7
Establish Quadrature Manually