![Keysight Technologies N5511A User Manual Download Page 288](http://html1.mh-extra.com/html/keysight-technologies/n5511a/n5511a_user-manual_1962254288.webp)
288 N5511A Phase Noise Test System User’s Guide
Baseband Noise Measurement Example
Baseband Noise with Test Set Measurement Example
lists the parameter data used for this measurement example.
Table 13-1
Parameter data for the baseband using a test set measurement
Step
Parameters
Data
1
Type and Range Tab
Measurement Type
Baseband Noise (using a test set)
Start Frequency
10 Hz
Stop Frequency
160 E + 6 Hz
Averages
1
Cross-correlations
1
Channel Setup
Dual
Quality
Fast
2
Cal Tab
Gain preceding noise input
0 dB
3
Block Diagram Tab
Noise Source
Test Set Noise Input
4
Test Set Tab
Input Attenuation
0 dB
LNA Low Pass Filter
20 MHz (Auto checked)
LNA Gain
Auto Gain (Minimum Auto Gain –14 dB)
DC Block
Not checked
PLL Integrator Attenuation
0 dBm
5
Graph Tab
Title
Baseband using the N5500A test set
Graph Type
Baseband Noise (dBV)
X Scale Minimum
1 Hz
X Scale Maximum
160 E + 6 Hz
Y Scale Minimum
-50 dBV/Hz
Y Scale Maximum
–210 dBV/Hz
Normalize trace data to a:
1 Hz bandwidth
Scale trace data to a new carrier frequency of:
1 times the current carrier frequency
Shift trace data DOWN by
0 dB
Trace Smoothing Amount
0
Power present at input of DUT
0 dB
Summary of Contents for N5511A
Page 1: ...User s Guide Keysight N5511A Phase Noise Test System...
Page 4: ...4...
Page 11: ...Contents N5511A Phase Noise Test System User s Guide 11...
Page 12: ...12 N5511A Phase Noise Test System User s Guide Contents...
Page 38: ...38 N5511A Phase Noise Test System User s Guide Phase Noise Basics What is Phase Noise...