Absolute Measurement Fundamentals
Calibration
Estimate the system's capture range (using the VCO source parameters
entered for this measurement). The estimated VCO tuning constant must be
accurate within a factor of 2. A procedure for
“Estimating the Tuning Constant”
is located in this chapter.
If you are able to locate the beat note, but it distorts and then disappears as you adjust it towards
0 Hz, your sources are injection locking to each other. Set the beat note to the lowest frequency
possible before injection locking occurs and then refer to the
“Minimizing Injection Locking”
section of this chapter for recommended actions.
If you are not able to tune the beat note to within the capture range due to frequency drift, refer to
the
section of this chapter for information about measuring
drifting signals.