134
Keysight M8131A 16/32 GSa/s Digitizer User’s Guide
4
Remote Programming
:WAVeform:XORigin?
Command
WAVeform:XORigin? CHANnel<N>
Description
This query returns the y-origin voltage value for the channel given by the
first parameter. The voltage value returned is the voltage value
represented by the waveform data digital code zero.
<N> An integer, 1-4.
Example
:WAVeform:YINCrement? CHAN1 -> 5.00000000000000E-04
:WAVeform:YINCrement?
Command
:WAVeform:YINCrement? CHANnel<N>
Description
This query returns the y-increment voltage for the channel given by the
first parameter. This voltage value is the voltage difference between two
adjacent waveform data digital codes.
<N> An integer, 1-4.
Example
WAVeform:XORigin? CHAN1 -> 0.0000000000
:WAVeform:YORigin?
Command
WAVeform:YORigin? CHANnel<N>
Description
This query returns the y-origin voltage value for the channel given by the
first parameter. The voltage value returned is the voltage value
represented by the waveform data digital code zero.
<N> An integer, 1-4.
Example
WAVeform:YORigin? CHAN1 -> 0.0000000000