150
Keysight M8131A 16/32 GSa/s Digitizer User’s Guide
4
Remote Programming
: TRIGger:EDGE:SLOPe
Command
:TRIGger:EDGE:SLOPe {POSitive | NEGative}
Description
The command sets the slope of the trigger source. The slope setting is
always the same for all trigger sources.
POSitive Trigger on a rising edge.
NEGative Trigger on a falling edge.
Query
:TRIGger:EDGE:SLOPe?
The query returns the current setting for the trigger source.
Returned Format
[:TRIGger:EDGE:SLOPe] {EXTernal | CHANnel<N>}<NL>
:TRIGger:LEVel:EXTernal
Command
:TRIGger:LEVel:EXTernal <level>
Description
The command specifies the trigger level when the external trigger input is
used as trigger source.
<level> The trigger level in Volts as a floating point number.
Query
:TRIGger:LEVel:EXTernal?
The query returns the current trigger level.
Returned Format
[:TRIGger:LEVel:EXTernal] <level><NL>
:TRIGger:LEVel[:CHANnel]
Command
:TRIGger:LEVel[:CHANnel] CHANnel<N>, <level>
Description
The command specifies the trigger level when an analog input channel is
used as trigger source.
<N> An integer, 1-4.
<level> The trigger level in Volts as a floating point number.
Query
:TRIGger:LEVel[:CHANnel]? CHANnel<N>
The query returns the current trigger level.
Returned Format
[:TRIGger:LEVel[:CHANnel]] <level><NL>