Keysight M8131A 16/32 GSa/s Digitizer User’s Guide
143
Remote Programming
4
IBANd
In-band
<rx_flow_control>
Receive flow control
NONE
IBANd In-band
<options>
String to pass additional instrument-specific settings. For
future expansion. Currently not used.
Query
:ODI:PORT<N>:ACTivate?
Description
When the port is active, this query returns the parameter values used in
the command to activate the port.
:ODI:PORT<N>:DEACtivate
Command
:ODI:PORT<N>:DEACtivate
Description
The command switches off the optical port.
<N>
An integer to select the ODI port, 1-4.
Example
:ODI:PORT1:DEAC
:ODI:PORT<N>:CSTatus?
Query
:ODI:PORT<N>:CSTatus?
Description
The query returns the communication status of an optical port.
<N>
An integer to select the ODI port, 1-4.
Returned Format
[:ODI:PORT<N>:CSTatus] <status><NL>
<status>
32-bit integer, meaning of the status bits is described in