40
Keysight M8131A 16/32 GSa/s Digitizer User’s Guide
2
M8131A Software Installation
Power and Ventilation Requirements
For power and ventilation requirements, refer to:
•
http://www.keysight.com/find/M9514A
for 14-slot chassis related
documentation.
•
http://www.keysight.com/find/M9505A
for 5-slot chassis related
documentation.
•
http://www.keysight.com/find/M9502A
for 2-slot chassis related
documentation.
Thermal Protection
Overheating Detection
The instrument monitors its internal temperature. If the temperature
exceeds approximately 80°C the power supply is switched off. The
instrument will not turn on automatically if the temperature is decreasing
again.
Fan Failure
If a fan is broken or prevented from operating by a blockage the
temperature will increase. When the temperature exceeds approximately
80°C the overheating detection switches off the instrument for safety
reasons. For reliability it is recommended to send instruments with broken
or defective fans immediately to Keysight Service for repair.
Battery
The M8131A does not have a battery.
Operating Environment
For details on operative environment for M8131A, refer to the section