![Keysight Technologies M8131A User Manual Download Page 101](http://html1.mh-extra.com/html/keysight-technologies/m8131a/m8131a_user-manual_1961999101.webp)
Keysight M8131A 16/32 GSa/s Digitizer User’s Guide
101
Remote Programming
4
:CALibrate:ADC:INFormation?
Query
:CALibrate:ADC:INFormation?
Description
The query returns information about the last calibration of the ADC
module.
Returned Format
[:CALibrate:ADC:INFormation] <information_string><NL>
<information_string>
Contains information about type (“User” or
“Factory”) and date of last calibration.
Example
Query
:CALibrate:ADC:INFormation?
“M8131A Setup:
Mode: 16GHz on 4 channels
Calibration Loaded:
1(User): 2018-03-09 17:33:44”
:CALibrate:ADC:RECommended?
Query
:CALibrate:ADC:RECommended?
Parameters
None
Description
This query returns a Boolean value, indicating if an ADC calibration is
recommended.
Return Format
[:CALibrate:ADC:RECommended] <recommended><NL>
Example
:CAL:ADC:REC?
0
:CALibrate:ADC:DATE[:RECommended]?
Query
:CALibrate:ADC:DATE[:RECommended]?
Parameters
None
Description
This query returns the date when the next ADC calibration is
recommended.
Return Format
[:CALibrate:ADC:DATE:RECommended] <date><NL>
Example
:CAL:ADC:DATE?
“2020-07-01”