Keysight M8131A 16/32 GSa/s Digitizer User’s Guide
103
Remote Programming
4
:CALibrate:ADC:SAVE
Command
:CALibrate:ADC:SAVE <file_name>
Description
The command saves the calibration data of the ADC module to a file.
<file_name>
A quoted ASCII string with a maximum of 254 characters
including the entire path name, if used. The filename assumes the present
working directory, if a path does not precede the file name.
:CALibrate:ADC:LOAD
Command
:CALibrate:ADC:LOAD <file_name>
Description
The command loads the calibration data of the ADC module from a file.
<file_name>
A quoted ASCII string with a maximum of 254 characters
including the entire path name, if used. The filename assumes the present
working directory, if a path does not precede the file name.