Keysight M8131A 16/32 GSa/s Digitizer User’s Guide
95
Remote Programming
4
ADC Calibration Run Status Subsystem
The following SCPI commands and queries are supported:
:STATus:OPERation:CALibration:ADC[:EVENt]?
:STATus:OPERation:CALibration:ADC:CONDition?
:STATus:OPERation:CALibration:ADC:ENABle[?]
:STATus:OPERation:CALibration:ADC:NTRansition[?]
:STATus:OPERation:CALibration:ADC:PTRansition[?]
Table 14
Run status register
Bit Number
Decimal Value
Definition
0
Calibration Run Status
1
Indicates that an ADC calibration is running
1
Calibration Step Running
2
Indicates that an ADC calibration step is running
NOTE
In modules with differential inputs, only bit 0 is important. It is set while
the calibration is active and is cleared when the calibration is finished.
In modules with single ended inputs, where an external calibration probe
must be connected for the calibration, this is different. These modules
are calibrated channel by channel and for each step the calibration probe
must be connected to the next channel.
Bit 0 is set when the calibration for the first channel is started and
cleared when calibration for the last channel is finished.
Bit 1 is set while the calibration for a single channel is running. A ‘0’
indicates that the calibration probe can be connected to the next
channel and the :CALibrate:ADC:NEXT SCPI command can be issued
again.