Measurements
14
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
191
Where x
i
= value at
i
th point being measured, n = number of points in
measurement interval.
The Full Screen measurement interval variation measures the value on all
displayed data points.
The N Cycles measurement interval variation measures the value on an integral
number of periods of the displayed signal. If less than three edges are present, the
measurement shows "No edges".
The X cursors show the interval of the waveform being measured.
AC RMS
AC RMS is the root-mean-square value of the waveform, with the DC component
removed. It is useful, for example, for measuring power supply noise.
The N Cycles measurement interval measures the value on an integral number of
periods of the displayed signal. If less than three edges are present, the
measurement shows "No edges".
The X cursors show the interval of the waveform being measured.
The Full Screen (Std Deviation) measurement interval variation is an RMS
measurement across the full screen with the DC component removed. It shows the
standard deviation of the displayed voltage values.
The standard deviation of a measurement is the amount that a measurement
varies from the mean value. The Mean value of a measurement is the statistical
average of the measurement.
The following figure graphically shows the mean and standard deviation. Standard
deviation is represented by the Greek letter sigma:
σ
. For a Gaussian distribution,
two sigma (± 1
σ
) from the mean, is where 68.3 percent of the measurement results
reside. Six sigma (± 3
σ
) from is where 99.7 percent of the measurement results
reside.
Summary of Contents for DSO-X 2002A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 22: ...22 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 70: ...70 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 84: ...84 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 106: ...106 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 112: ...112 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 146: ...146 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 154: ...154 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 180: ...180 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 216: ...216 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 228: ...228 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 246: ...246 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 266: ...266 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Utility Settings...
Page 352: ...352 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide Index...