Math Waveforms
4
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
81
• Adjust frequency span for better cursor placement.
• Return to the Cursors Menu to fine tune the X cursor.
For more information on the use of FFTs please refer to Keysight Application Note
243,
The Fundamentals of Signal Analysis
at
http://literature.cdn.keysight.com/litweb/pdf/5952-8898E.pdf
. Additional
information can be obtained from Chapter 4 of the book
Spectrum and Network
Measurements
by Robert A. Witte.
FFT Units
0 dBV is the amplitude of a 1 Vrms sinusoid. When the FFT source is channel 1 or
channel 2 (or channel 3 or 4 on 4-channel models), FFT units will be displayed in
dBV when channel units is set to Volts and channel impedance is set to 1 M
Ω
.
FFT units will be displayed in dBm when channel units is set to Volts and channel
impedance is set to 50
Ω
.
FFT units will be displayed as dB for all other FFT sources or when a source
channel's units has been set to Amps.
FFT DC Value
The FFT computation produces a DC value that is incorrect. It does not take the
offset at center screen into account. The DC value is not corrected in order to
accurately represent frequency components near DC.
FFT Aliasing
When using FFTs, it is important to be aware of frequency aliasing. This requires
that the operator have some knowledge as to what the frequency domain should
contain, and also consider the sampling rate, frequency span, and oscilloscope
vertical bandwidth when making FFT measurements. The FFT resolution (the
quotient of the sampling rate and the number of FFT points) is displayed directly
above the softkeys when the FFT Menu is displayed.
Summary of Contents for DSO-X 2002A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 22: ...22 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 70: ...70 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 84: ...84 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 106: ...106 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 112: ...112 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 146: ...146 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 154: ...154 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 180: ...180 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 216: ...216 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 228: ...228 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 246: ...246 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 266: ...266 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Utility Settings...
Page 352: ...352 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide Index...