Triggers
10
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
127
The trigger level for a selected digital channel is set using the threshold menu in
the Digital Channel Menu. Press the
[Digital]
key on the front panel, then press the
Thresholds
softkey to set the threshold level (TTL, CMOS, ECL, or user defined) for
the selected digital channel group. The threshold value is displayed in the
upper-right corner of the display.
The line trigger level is not adjustable. This trigger is synchronized with the power
line supplied to the oscilloscope.
Forcing a Trigger
The
[Force Trigger]
key causes a trigger (on anything) and displays the acquisition.
This key is useful in the Normal trigger mode where acquisitions are made only
when the trigger condition is met. In this mode, if no triggers are occurring (that is,
the "Trig'd?" indicator is displayed), you can press
[Force Trigger]
to force a trigger
and see what the input signals look like.
In the Auto trigger mode, when the trigger condition is not met, triggers are forced
and the "Auto?" indicator is displayed.
Edge Trigger
The Edge trigger type identifies a trigger by looking for a specified edge (slope)
and voltage level on a waveform. You can define the trigger source and slope in
this menu. The trigger type, source, and level (if applicable) are displayed in the
upper-right corner of the display.
1
On the front panel, in the Trigger section, press the
[Trigger]
key.
2
In the Trigger Menu, press the
Trigger
softkey, and use the Entry knob to select
Edge
.
3
Select the trigger source:
•
Analog channel,
1
to the number of channels
NOTE
You can also change the trigger level of all channels by pressing
[Analyze] > Features
and
then selecting
Trigger Levels
.
Summary of Contents for DSO-X 2002A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 22: ...22 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 70: ...70 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 84: ...84 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 106: ...106 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 112: ...112 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 146: ...146 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 154: ...154 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 180: ...180 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 216: ...216 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 228: ...228 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 246: ...246 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 266: ...266 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Utility Settings...
Page 352: ...352 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide Index...