Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
4
Math Waveforms
FFT Measurement Hints
The number of points acquired for the FFT record can be up to 65,536, and when
frequency span is at maximum, all points are displayed. Once the FFT spectrum is
displayed, the frequency span and center frequency controls are used much like
the controls of a spectrum analyzer to examine the frequency of interest in greater
detail. Place the desired part of the waveform at the center of the screen and
decrease frequency span to increase the display resolution. As frequency span is
decreased, the number of points shown is reduced, and the display is magnified.
While the FFT spectrum is displayed, use the
[Math]
and
[Cursors]
keys to switch
between measurement functions and frequency domain controls in FFT Menu.
Decreasing the effective sampling rate by selecting a greater time/div setting will
increase the low frequency resolution of the FFT display and also increase the
chance that an alias will be displayed. The resolution of the FFT is the effective
sample rate divided by the number of points in the FFT. The actual resolution of
the display will not be this fine as the shape of the window will be the actual
limiting factor in the FFTs ability to resolve two closely space frequencies. A good
way to test the ability of the FFT to resolve two closely spaced frequencies is to
examine the sidebands of an amplitude modulated sine wave.
For the best vertical accuracy on peak measurements:
• Make sure the probe attenuation is set correctly. The probe attenuation is set
from the Channel Menu if the operand is a channel.
• Set the source sensitivity so that the input signal is near full screen, but not
clipped.
• Use the Flat Top window.
• Set the FFT sensitivity to a sensitive range, such as 2 dB/division.
For best frequency accuracy on peaks:
• Use the Hanning window.
• Use Cursors to place an X cursor on the frequency of interest.
NOTE
FFT Resolution
The FFT resolution is the quotient of the sampling rate and the number of FFT points (f
S
/N).
With a fixed number of FFT points (up to 65,536), the lower the sampling rate, the better the
resolution.
Summary of Contents for DSO-X 2002A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 22: ...22 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 70: ...70 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 84: ...84 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 106: ...106 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 112: ...112 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 146: ...146 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 154: ...154 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 180: ...180 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 216: ...216 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 228: ...228 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 246: ...246 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 266: ...266 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Utility Settings...
Page 352: ...352 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide Index...