200
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
14
Measurements
4
Press the
Lower
softkey; then, turn the Entry knob to set the lower measurement
threshold value.
Increasing the lower value beyond the set middle value will automatically
increase the middle value to be more than the lower value. The default lower
threshold is 10% or 800 mV.
If threshold
Type
is set to
%
, the lower threshold value can be set from 5% to
93%.
5
Press the
Middle
softkey; then, turn the Entry knob to set the middle
measurement threshold value.
The middle value is bounded by the values set for lower and upper thresholds.
The default middle threshold is 50% or 1.20 V.
•
If threshold
Type
is set to
%
, the middle threshold value can be set from 6% to
94%.
6
Press the
Upper
softkey; then, turn the Entry knob to set the upper
measurement threshold value.
Decreasing the upper value below the set middle value will automatically
decrease the middle value to be less than the upper value. The default upper
threshold is 90% or 1.50 V.
•
If threshold
Type
is set to
%
, the upper threshold value can be set from 7% to
95%.
Measurement Window with Zoom Display
When the zoomed time base is displayed, you can choose whether measurements
are made in the Main window portion of the display or the Zoom window portion of
the display.
1
Press the
[Meas]
key.
2
In the Measurement Menu, press the
Settings
softkey.
3
In the Measurement Settings Menu, press the
Meas Window
softkey; then, turn
the Entry knob to select from:
•
Auto Select
— The measurement is attempted in the lower, Zoom window; if it
cannot be made there, the upper, Main window is used.
•
Main
— The measurement window is the upper, Main window.
•
Zoom
— The measurement window is the lower, Zoom window.
Summary of Contents for DSO-X 2002A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 22: ...22 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 70: ...70 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 84: ...84 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 106: ...106 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 112: ...112 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 146: ...146 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 154: ...154 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 180: ...180 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 216: ...216 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 228: ...228 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 246: ...246 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 266: ...266 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Utility Settings...
Page 352: ...352 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide Index...