Waveform Generator
17
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
219
For all waveform types, the output amplitude, into 50
Ω
, can be adjusted from
10 mVpp to 2.5 Vpp (or from 20 mVpp to 5 Vpp into an open-circuit load).
Pressing a signal parameter softkey can open a menu for selecting the type of
adjustment. For example, you can choose to enter amplitude and offset values,
or you can choose to enter high-level and low-level values. Or, you can choose
to enter frequency values or period values. Keep pressing the softkey to select
the type of adjustment. Turn the Entry knob to adjust the value.
Notice that you can select between coarse and fine adjustments for frequency,
period, and width. Also, pushing the Entry knob is a quick way to toggle
between coarse and fine adjustments.
Waveform
Type
Characteristics
Sine
Use the
Frequency/Frequency Fine/Period/Period Fine
,
Amplitude
/
High-Level
, and
Offset
/
Low-Level
softkeys to set the sine signal parameters.
The frequency can be adjusted from 100 mHz to 20 MHz.
Square
Use the
Frequency/Frequency Fine/Period/Period Fine
,
Amplitude
/
High-Level
,
Offset
/
Low-Level
, and
Duty Cycle
softkeys to set the square wave signal
parameters.
The frequency can be adjusted from 100 mHz to 10 MHz.
The duty cycle can be adjusted from 20% to 80%.
Ramp
Use the
Frequency/Frequency Fine/Period/Period Fine
,
Amplitude
/
High-Level
,
Offset
/
Low-Level
, and
Symmetry
softkeys to set the ramp signal parameters.
The frequency can be adjusted from 100 mHz to 100 kHz.
Symmetry represents the amount of time per cycle that the ramp waveform is rising
and can be adjusted from 0% to 100%.
Pulse
Use the
Frequency/Frequency Fine/Period/Period Fine
,
Amplitude
/
High-Level
,
Offset
/
Low-Level
, and
Wid th/Wid th Fine
softkeys to set the pulse signal
parameters.
The frequency can be adjusted from 100 mHz to 10 MHz.
The pulse width can be adjusted from 20 ns to the period minus 20 ns.
DC
Use the
Offset
softkey to set the DC level.
Noise
Use the
Amplitude
/
High-Level
and
Offset
/
Low-Level
to set the noise signal
parameters.
Summary of Contents for DSO-X 2002A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 22: ...22 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 70: ...70 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 84: ...84 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 106: ...106 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 112: ...112 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 146: ...146 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 154: ...154 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 180: ...180 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 216: ...216 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 228: ...228 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 246: ...246 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 266: ...266 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Utility Settings...
Page 352: ...352 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide Index...