CALibrate:STATe ON|1|OFF|0 [,<password>]
CALibrate:STATe?
Enables or disables calibration mode. Calibration mode must be enabled for the instrument to accept
any calibration commands.
Parameter
Typical Return
ON|1|OFF|0 , default OFF
0 (OFF) or 1 (ON)
<
password
> a numeric value up to 15 digits (none)
Disable calibration: CAL:STAT OFF
Enable calibration: CAL:STAT ON ,"abc123"
The <
password
> is required if it has been set. The factory default is no password, but if one has been
set, you must send CALibrate:STATe ON,12345 (replacing 12345 with your instrument's password).
You cannot enable front-panel calibration and SCPI calibration simultaneously.
CALibrate:VOLTage[:LEVel] <
value
>
CALibrate:VOLTage:OFFset <
value
>
Selects the voltage range to calibrate.
Parameter
Typical Return
Either 219 or 438 (for DC OFFset) or 155 or 310 (for AC LEVel). (none)
Calibrate the voltage of the 310 V range: CAL:VOLT:LEV 310
The
must be ON to execute this command.
The CALibrate:VOLTage[:LEVel] command is for AC voltage, and the CALibrate:VOLTage:OFFSet
command is for DC voltage.
Do not use unit suffixes, such as V or mV, with this command.
3 SCPI Programming Reference
120
Keysight AC6800B Series Operating and Service Guide
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...