Selecting the Output Voltage Programming Source and Output Mode
Select the Output Voltage Programming Source
The EXT-AC and EXT-DC modes require the optional analog interface board (option
AC68BALGU). You cannot use the analog output interface board with the GPIB interface
because both boards use the same slot.
Without the analog interface board, only the Internal option is available. This controls the output
voltage by the instrument. The Internal option is also selected by
.
EXT_AC configures the instrument to behave as a variable gain amplifier controlled by an external DC
level. The internal frequency control remains in effect.
EXT_DC mode configures the instrument to behave as a fixed gain power amplifier. Internal controls
for voltage and frequency are disabled.
You cannot set the voltage limit when the instrument is being controlled by external
analog signals. An excessive external voltage may damage the load.
Front Panel:
1.
Select
[Menu] > Output > Voltage > Prog
.
2.
Select
Internal
,
EXT_AC
, or
EXT_DC
and press
[Select]
.
SCPI:
Select the Output Mode
You can switch the output among the five modes shown below when the output is turned off.
The EXT-AC and EXT-DC modes require the optional analog interface board. You cannot
use the analog output interface board with the GPIB interface because both boards use
the same slot.
Output Mode Description
AC Mode
Produce AC output
DC mode
Produce DC output
AC+DC mode Superimpose DC voltage on the AC output
EXT-AC mode Output sine waves using external DC signals
EXT-DC mode Amplify and output the waveform applied externally
Keysight AC6800B Series Operating and Service Guide
71
2 User Information
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...