STATus:QUEStionable:NTRansition <
value
>
STATus:QUEStionable:NTRansition?
STATus:QUEStionable:PTRansition <
value
>
STATus:QUEStionable:PTRansition?
Sets and queries the value of the
(Negative-Transition) and
(Positive-Transition) registers.
These registers serve as a polarity filter between the Questionable Condition and Questionable Event
registers.
When a bit in the NTR register is 1, then a 1-to-0 transition of the corresponding bit in the
Questionable Condition register causes that bit in the Questionable Event register to be set.
When a bit in the PTR register is 1, then a 0-to-1 transition of the corresponding bit in the Questionable
Condition register causes that bit in the Questionable Event register to be set.
The
command sets all bits in the PTR registers and clears all bits in the NTR registers.
Parameter
Typical Return
Decimal sum of the bits in the register, default 0. For example, to enable bit 2 (value 4), bit 3 (value 8), and
bit 7 (value 128), use decimal sum 140 (4 + 8 + 128). Default 0.
<
bit value
>
Enable bits 3 and 4 in the questionable PTR register: STAT:QUES:PTR 24
If the same bits in both NTR and PTR registers are set to 1, then any transition of that bit at the
Questionable Condition register sets the corresponding bit in the Questionable Event register.
If the same bits in both NTR and PTR registers are set to 0, then no transition of that bit at the
Questionable Condition register can set the corresponding bit in the Questionable Event register.
The value returned is the binary-weighted sum of all bits set in the register.
3 SCPI Programming Reference
150
Keysight AC6800B Series Operating and Service Guide
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...