End Calibration
This section describes the final steps in the calibration process.
Storing calibration constants overwrites the existing ones in nonvolatile memory. If you
do not save the calibration data, the new calibration constants will only be used until you
exit the calibration state or cycle power. For the front panel, leaving the calibration state
is defined as exiting the menu levels at or below [Menu] > System > Admin > Cal. For
SCPI, leaving the calibration state is accomplished by CAL:STAT OFF.
If you are not sure you want to permanently store the new constants, do not save the
data when you exit the calibration mode.
Step
Description
Front Panel
SCPI
1
Enter the calibration
date.
[Menu] > System > Admin > Cal > Date
Enter the calibration date in the
Data
field. Press
[Select]
or
[Enter]
to finish.
CAL:DATE "<
string
>"
2
Save calibration data.
[Menu] > System > Admin > Cal > Save
CAL:SAVE
3
Log out.
[Menu] > System > Admin > Logout
CAL:STAT OFF
4 Calibration, Verification, and Service
176
Keysight AC6800B Series Operating and Service Guide