The Biofuge primo
11
The "Easycontrol" user interface
Function Feature
lid opening
electromagnetic release via "open lid“ key when mains supply OK
(unlocking in case of power failure: see chapter "Troubleshooting")
start
start key (
)
stop
stop key (
)
short-term acceleration
"quick run" key ( ): short-term run when permanently pressed; stop upon key
release
acceleration/braking profiles
1 = slow acceleration and slowest braking curve (2)
2 - 9 = fast acceleration and 8 different braking profiles
speed selection
adjustable in steps of 10 min
-1
within the range of 300 min
-1
to 15000 min
-1
RCF display
upon actuating the switchover key, the current RCF value is displayed
run time selection
adjustable in minutes from 1 min to 9 h 59 min; "hLd" mode: permanent operation
run time display in "quick run" mode
between 1 s and 60 s in seconds steps, above in minutes
end of run
speed display reads "End"