Troubleshooting
53
Error
Behavior of the
centrifuge
Possible cause(s) and measures to be taken
E-24
NV-RAM 2 absent
E-25
Rotor stops without brak-
ing to standstill.
Start without rotor.
1. Turn instrument off and on again.
2. Open the instrument by pressing the "Open lid" key ..
3. Check whether the rotor is loaded and placed correctly.
4. Check whether a broken vessel or a damaged rotor was re-
sponsible for actuating the unbalance switch.
If the error persists, call Service.