Accessories
22
Highconic
®
- rotor
You must always equip all sample
locations with adapters!
In case longer tubes shall be centrifuged
precluding the complete locking of the rotor
lid, it is allowed to operate the rotor without
lid, up to a maximum speed of 4000 rpm.
Above 4000 rpm the rotor may not
be operated without lid in any case.
Otherwise the rotor could be de-
stroyed!
Drum rotor
You must always equip all sample
locations with adapters or sample
racks!