Troubleshooting
52
Error
Behavior of the
centrifuge
Possible cause(s) and measures to be taken
E-10
NV-RAM; error in program memory
Switch the instrument off and on again. If the problem persists, call
Service.
E-14
Instrument does not
starts or brakes to stand-
still.
No rotor present or rotor identification impossible
A) Check whether a rotor is inserted.
B) Following a brief power failure, the rotor could not be identified.
Switch the instrument off and on again using the mains switch.
C) Phases at the motor mixed up.
E-15
Rotor stops without brak-
ing to standstill.
Instrument cannot be
operated.
Check sum in NV-RAM wrong.
E-17
Lid does not open.
Lid blocked or jammed.
Press the front part of the lid centrally down once, and press the
"open lid" key anew.
Otherwise see "Emergency lid release"
E-19
Wrong NV-RAM or keyboard
E-22
NV-RAM parameter incompatible with processor