Maintenance and care
41
Rotor and rotor chamber must be treated with a neu-
tral, universal disinfectant. Best suited for this purpose
are disinfectant sprays, ensuring that all rotor and ac-
cessory surfaces are covered evenly.
•
Please use 70% ethanol for disinfection.
Please note the safety measures and
handling hints when applying these
substances!
For other disinfectants please consult KENDRO
Services!
•
You may disinfect the rotor and the accessories as
described in the following section. Be sure to follow
the pertinent safety procedures for handling infec-
tious material.
1. Pull mains plug.
2. Unscrew the rotor chuck.
3. Grab the rotor with both hands and pull it perpen-
dicularly off the drive shaft.
4. Remove the centrifuge tubes and adapters, and
disinfect them or dispose of them as necessary.
5. Treat the rotor and the rotor lid according to the
instructions given for the disinfectant (soaking in
liquid or spraying). You must strictly observe the
specified action times!
6. Turn the rotor head down and drain off the disinfec-
tant. Thereafter thoroughly rinse rotor and lid with
water.
7. Dispose of the disinfectant according to valid regu-
lations.
8. Aluminum rotors have to be treated with anticorro-
sive protective oil subsequently.
Summary of Contents for SORVALL Biofuge primo
Page 1: ...SORVALL Biofuge primo OPERATING INSTRUCTIONS PN 00000 0...
Page 10: ...For your safety 6 Notes...
Page 12: ...The Biofuge primo 8 for your notes...
Page 42: ...Operation 38 for your notes...
Page 61: ...Appendix 57 Appendix...
Page 64: ...Appendix 60...
Page 71: ...Appendix 67 Autoclaving protocol Date Remark Operator Signature 1 2 3 4 5 6 7 8 9 10...
Page 72: ...Appendix 68...