Accessories
17
Table 1: Rotor program (2)
Rotor designation
microliter rotor
24 x 2 ml aluminium
microliter rotor
24 x 2 ml
Polypropylene
PCR-rotor
4 x PCR-Strip
order no.
75007593 75007587 75007569
maximum permissible load [ g ]
24 x 4
24 x 4
4 x 4
( 32 x 0,5 )
maximum speed n
max
[ min
-1
]
15 000
13 000
13 000
maximum RCF-value at n
max
21 882
16 060
12 846
( 11 524 )
2)
maximum radius [ cm ]
8.7
8.5
6,8
( 6.1)
2)
minimum radius [ cm ]
5.9
5.9
6,1
( 4.7 )
2)
angle [ ° ]
45
40
45
acceleration time [ s ]
40
15
15
braking time [ s ]
30
25
25
aerosol-tight
1)
yes
(reduced filling)
yes
(reduced filling)
yes
(reduced filling)
permissible temperature range
autoclavable (number of cycles)
–
121°C; (unlimited)
-4 °C to +40 °C
121°C, (10 cycles)
-4 °C to +40 °C
121°C, (10 cycles)
1) Checked by CAMR, Porton-Down, UK
2) The values relate to vessel places 4 and 5 in the PCR-Rotor