Maintenance and care
42
Disinfection with bleaching lye
These agents contain highly aggres-
sive hypochlorites and must not be
used with aluminum rotors!
To protect the rotor 75007587 and 75007569 as far as
possible you must take the following precautions:
1. Avoid high temperatures!
The bleaching solution and the rotor should not be
warmer than ca. 25 °C.
2. Do not let the bleaching solution act longer than
absolutely necessary!
3. After disinfection, rinse the rotor thoroughly with
distilled water and allow to dry.
Decontamination
For general radioactive decontamination, use a solu-
tion of equal parts of 70% ethanol, 10% SDS and wa-
ter. Follow this with ethanol rinses, then de-ionized
water rinses, and dry with a soft absorbent cloth.
Dispose of all washing solutions in appro-
priate radioactive waste containers!