![Kendro SORVALL Biofuge primo Operating Instructions Manual Download Page 20](http://html1.mh-extra.com/html/kendro/sorvall-biofuge-primo/sorvall-biofuge-primo_operating-instructions-manual_663812020.webp)
Accessories
16
Rotor program
Table 1: Rotor program (1)
Rotor designation
High
conic
®
rotor
6 x 50 ml
swinging bucket rotor
4 x 100 ml
drum rotor
order no.
75007588 75007591 75007595
buckets and caps see Table 2
maximum permissible load [ g ]
6 x 130
4 x 200
8 x 80
maximum speed n
max
[ min
-1
]
8500 4000 12
000
maximum RCF value at n
max
10 016
2522
14 005
maximum radius [ cm ]
12.4
14.1
8.7
minimum radius [ cm ]
6.0
5.0
3.8
angle [ ° ]
45
90
90 / 60
acceleration time [ s ]
55
25
40
braking time [ s ]
35
20
30
aerosol-tight
1)
yes
(reduced filling)
yes
no
permissible temperature range
autoclavable (number of cycles)
–
121 °C; (unlimited)
–
121 °C; (unlimited)
–
no
1) Checked by CAMR, Porton-Down, UK