Before use
13
Before use
Transport and installation
The centrifuge is delivered in an special box. Cut it
open and remove the protective material.
Damage to the centrifuge by jolting!
Transport the centrifuge only in the
upright position using the special box
provided with the instrument and se-
cure it properly. Place the centrifuge
carefully.
Proper location
The location for the centrifuge must meet the following
criteria:
•
A safety zone of at least 30 cm around the centri-
fuge must be maintained where neither persons nor
hazardous materials may be stationed during cen-
trifugation.
•
The substructure must be stable and resonance-
free. A good support is provided by a plane labora-
tory bench or a large laboratory carriage with lock-
able casters.
•
To ensure sufficient air circulation, a minimum dis-
tance from the wall of 10 cm at the back and of 15
cm on each side must be kept.
•
The centrifuge must be protected from heat and
direct sunshine.
•
The location must be well ventilated at all times.