Operation
19
Loading the hematocrit rotor
After filling and sealing the capillaries, place them into
the guide notches with the sealed ends pointing out-
wards.
The centrifuge must always be loaded
weight-symmetrically to avoid unbal-
ance, unsteady runs and subsequent
damage.
Inserting the hematocrit rotor
After checking for correct loading, you may insert the
rotor into the centrifuge.
Possible damage to drive and rotor!
Do not push down the rotor by force.
If the rotor is tilted, carefully remove
and reinsert it.
To put the rotor into position, proceed as follows:
1. Open the lid and make sure that the rotor chamber
is clean. Remove eventual dust, foreign material or
sample residues.
2. Turn the rotor so that the driving pin of the motor
shaft fits into the groove on the underside of the ro-
tor, and slide the rotor onto the drive shaft until the
rotor hub rests on the shaft (cf. Figure).
1
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
1
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
H
ER
AEU
S
INS
TR
UM
E
NT
S
Ca
t-N
o
1067
I
I
1
2
0
0
0
/
R
PM
I
I
H
ER
AEU
S
INS
TR
UM
E
NT
S
Ca
t-N
o
1067
I
I
1
2
0
0
0
/
R
PM
I
I
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
Befestigungsmutter
Summary of Contents for Heraeus Biofuge haemo
Page 1: ...Instructions for use Biofuge haemo ...
Page 3: ...Biofuge haemo set set 1 2 0 0 0 1 0 speed time lid quick run start stop INSTRUMENTS ...
Page 10: ...For your safety 6 for your notes ...
Page 18: ...Accessories 14 for your notes ...
Page 20: ...Before use 16 for your notes ...
Page 41: ...Operation 37 for your notes ...
Page 51: ...Troubleshooting 47 ...